Comparative localized linear accuracy of small-field cone-beam CT and multislice CT for alveolar bone measurements.

Loubele M, Van Assche N, Carpentier K, Maes F, Jacobs R, van Steenberghe D, Suetens P

Medical Image Computing (ESAT/PSI), Faculty of Engineering, Katholieke Universiteit Leuven, Leuven, Belgium. miet.loubele@uz.kuleuven.ac.be

OBJECTIVES: To compare the accuracy of cone-beam computerized tomography (CBCT) and multislice CT (MSCT) for linear jaw bone measurements. STUDY DESIGN: An ex vivo formalin-fixed human maxilla was imaged with both CBCT (Accuitomo 3D; Morita, Kyoto, Japan) and MSCT (4-slice Somatom VolumeZoom and 16-slice Somatom Sensation 16; Siemens, Erlangen, Germany). The MSCT images were reconstructed using different reconstruction filters to optimize bone visualization (U70u and U90u for VolumeZoom, H30s and H60s for Sensation 16). Before scanning, triplets of small gutta-percha markers were glued onto the soft tissues overlying the maxillary bone on the top and on both sides of the alveolar ridge to define a set of reproducible linear measurements in 11 planes. Image measurements were performed by 2 observers. The gold standard was determined by means of physical measurements with a caliper by 3 observers. RESULTS: The accuracy of the linear measurements was 0.35 +/- 1.31 mm (U70u) and 0.06 +/- 1.23 mm (U90u) for the Somatom VolumeZoom, 0.24 +/- 1.20 mm (H60s) and 0.54 +/- 1.14 mm (H30s) for the Sensation 16, and -0.09 +/- 1.64 mm for the Accuitomo 3D. Statistical analysis with 2-way analysis of variance showed no significant inter- or intraobserver disagreement for the physical or the radiologic measurements. There was also no significant difference for the measurements on the different reconstruction filters. CONCLUSION: Both CBCT and MSCT yield submillimeter accuracy for linear measurements on an ex vivo specimen.

Published 10 March 2008 in Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 105(4): 512-8.
Full-text of this article is available online (may require subscription).

Place a permanent text-link or advertisement here for just US$15.

© 2005-2008 Biostatistics Research Today. All Rights Reserved.



Biostatistics Research Today Archive:

Volume 1 (2005)
  Issue 1 (September)
  Issue 2 (October)
  Issue 3 (November)
  Issue 4 (December)

Volume 2 (2006)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)
  Issue 10 (October)
  Issue 11 (November)
  Issue 12 (December)

Volume 3 (2007)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)
  Issue 10 (October)
  Issue 11 (November)
  Issue 12 (December)

Volume 4 (2008)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)



Biostatistics Books

Blueprints Obstetrics and Gynecology (Blueprints Series)

Blueprints Obstetrics and Gynecology (Blueprints Series)